1.What is the specific scope of this atomizer?
|
This product is suitable for children and adults and can be used in hospitals.
Atomization treatment of various types of aerosolized drugs. Children and needs Adults with special care should be under the supervision of the guardian use Note: This product is a tool used for aerosol therapy, itself Does not contain aerosolized drugs. The type and amount of the drug Usage, please follow the doctor's advice. |
2.What is the specific method of operation?
|
1). Open the liquid medicine cup and inject the liquid medicine (please follow the doctor's instructions for atomization)
2). Attach the mask to the spray port (children or adults over 5 years old) Can use the mouthpiece); 3). Press the front of the fuselage to open the key, then the atomizer will start Spray, green light flashes;
4) Place the mask on the nose and mouth to make it fit completely.
Slow and deep breathing begins to atomize. Will automatically close after completion machine. If you need to pause, you can switch it back on. |
3.How to clean and disinfect?
|
1). cleaning: liquid cup, mask, mouthpiece can be soaked in clean water
Allow to dry naturally after washing. Do not rinse directly under the tap Liquid medicine cup; it is strictly forbidden to use hot water above 60° 2). Disinfection: The liquid cup can be submerged in 75% of medical alcohol. Dry out after the clock; face mask, mouthpiece can use 75% of the doctor Wipe with alcohol for more than 3 minutes, then dry Please keep it in a clean and dry place after drying thoroughly. more Please refer to the product manual for more details. |
4.What if I can't boot?
|
Please ensure that the product is powered properly.
1). When using battery power, please make sure the battery has power and press The positive and negative poles shown in the tank are correctly loaded. Open when the battery is too low Machine, the indicator light will be red, indicating that the battery needs to be more change |
5.Normal start, green light is on, but no fog is going on?
|
1). There is no liquid in the liquid medicine cup: please inject the liquid correctly and then turn it on.
Atomization. 2). Please make sure the liquid cup is in place: align the liquid cup to the host Chute, fully pushed to the body mount until you hear “kada ” sound completed installation. The liquid cup is in contact with the host electrode Sufficient to ensure normal spray. 3). atomization net blockage: put about 2m in the liquid medicine cup
At 60 ° C warm water, add a few drops of white vinegar, long press for 5 seconds to open
The key enters the self-cleaning mode, waiting for the atomization to complete automatically. Machine, that is, complete maintenance. 4). Please make sure the host electrode is clean: wipe off the liquid on the electrode Or water stains, use alcohol to remove dirt. 5). Due to damage to the atomized mesh or other reasons, it still cannot be solved. Please consult the official customer service. |
6.After the normal boot, how can it not continue to fog?
|
1). Please make sure the liquid cup is in place.
2). The atomizer only supports medical atomized liquid and saline mist. Please do not use pure water, distilled water, etc. Liquid. There are few conductive particles in this type of liquid, nebulizer Cannot be recognized properly. 3). Please ensure that the liquid is in full contact with the atomizing mesh. Can be shaken gently Shake the liquid bubble away, or adjust the holding angle to make the fog The spray nozzle of the chemicalizer is inclined downward by 45° to allow the liquid medicine and the spray port The mesh is in full contact. 4). If you still can't solve it, please consult the official customer service. |